  • Chlorine Analyzers FC500G

Chlorine Analyzers FC500G

2016-05-26 16:14:54

Description:

FC500G non-reagent free available chlorine analyzer adopts the polarographic method, using a rotating electrode to continuously measure online the concentration of free available chlorine. The FC500G is intended for measurements in the water distribution network, at pumping stations, water delivery points, schools, and public facilities.……

New Products

Honeywell 51307038-100 Power Control Module Assy

Honeywell 51307038-100 Power Control Module Assy. Honeywell Experion LX Series 8 C300 controller provides powerful and robust control for the Experion LX platform.

Honeywell 8937-HN2 Fibre Optic converter IO Module

Honeywell 8937-HN2 Multi-Mode, Fibre Optic converter IO Module. The Fiber Option Extender modules support the extension of the I/O link beyond the Series 8 cabinet with the S8 C300, enabling remote applications of the Series 8 I/O.

Honeywell 8939-HN Multi-Mode Fiber Optic Extender (IOTA)

Honeywell 8939-HN Multi-Mode Fiber Optic Extender IOTA. The Fiber Option Extender modules support the extension of the I/O link beyond the Series 8 cabinet with the S8 C300, enabling remote applications of the Series 8 I/O.

G.M. International X1-IS-AO-02-S I.S. SIL2 HART® Isolating Driver module

G.M. International X1-IS-AO-02-S I.S. SIL2 HART® Isolating Driver module. GMI Intrinsically Safe Isolators, Analog Output

Assortments

Description

FC500G

FC500G non-reagent free available chlorine analyzer adopts the polarographic method, using a rotating electrode to continuously measure online the concentration of free available chlorine.

The FC500G is intended for measurements in the water distribution network, at pumping stations, water delivery points, schools, and public facilities.

Features

  • Measuring range can be selected from 0 to 1/0 to 2/0 to 3mg/l and 0 to 5/0 to 10 mg/l.

  • Rotating electrode/ceramic bead cleaning has been improved to making cleaning automatic and continuous.

  • Easy-to-maintain (-clean) cell construction.

Specifications

Measured ObjectFree chlorine in tap water
Measuring SystemPolarographic method using rotating electrode
Measuring RangeAny of 0 to 1/0 to 2/0 to 3 mg/l and 0 to 5/0 to 10 mg/l certainly, use to air purge
Output Signal4 to 20mA DC (load resistance: Maximum 550 Ω)
Sample Solution ConditionTemperature: 0 to 40℃
pH: 6.5 to 7.5
Flow Rate: 0.1 to 2.5 l/min (at cell inlet)
Pressure: 1 to 30 kPa (without VALVE)
              1 to 150 kPa (with VALVE)
Conductivity: 100 to 300 μS/cm
SS (Suspended Solid): 10 mg/l or less
ElectrodeIndicating electrode: Rotating gold electrode
Reference electrode: Silver electrode (with Pt1000 RTD installed)
Electrode CleaningCeramic bead cleaning
Ambient Temperature-10 to 50℃
InstallationIndoor
Power Supply100 V AC/110 V AC / 200 V AC/220 V AC ±10 %, 50/60 Hz
WeightApprox. 6.5 kg

Performance

Linearity±5 % of full scale
Repeatability2 % of full scale
StabilityZero drift; ±1 % of full scale/month or less
Span drift; -10 % of full scale/month or less
Response timeApprox. 2 min (90 % response time)